What we do


We make scleral lenses of all kinds, but our main product is a range of gas permeable (RGP) scleral lenses made to specification from our fitting system.

 

Products

 

Innovative Sclerals Ltd is dedicated solely to the development and manufacture of scleral contact lenses (formally known as haptic lenses) and associated products and services. The main material used is clear gas permeable plastic formulated exclusively for Innovative Sclerals Ltd., which is 115Dk value measured by the 'Fatt' method. Clear PMMA (acrylic) is also used, mainly for diagnostic fitting lenses. These lenses are of the following types -

 

 

 

 

 

 

 

 

 

Standard Preformed RGP Scleral Lenses (lathe cut)

This is the main product, based on the 18 lens standard diagnostic scleral fitting system designed by Ken Pullum. They are normally 23 to 24mm total diameter with an optic diameter 10 to 12mm diameter. Lenses are normally made with the optic decentred approximately 1mm to improve centricity on the eye. These lenses will suit the majority of wearers for whom scleral lenses are indicated.
In addition a range of preformed 'Miniscleral' lenses have been successfully introduced which follow the same fitting principles as the standard range. These are 18mm total diameter with a co-axial optic zone. This range can be fitted both sealed and ventilated.  

 

 

 

 

 

 

Other Preformed RGP Scleral Lenses

Lenses can be supplied which vary one or more of the parameters on which the standard range is based - · Variation of limbal projection relative to apical projection where required to improve corneal clearance · Variation of scleral radius · Variation of lens diameter, including elliptical diameters · Variation of the decentration · Variation of the lens thickness in the scleral zone for certain indications, including heavy handling problems, and ptosis (Standard preformed lenses can, of course, incorporate variations to diameter, decentration and thickness.)

 

RGP Lenses from Impression

Innovative Sclerals Ltd have developed methods to manufacture non-coaxial scleral lenses in RGP. Thus it is now possible to accurately make scleral contact lenses based on a cast ('impression') of the patient's eye. This brings full circle the traditional method of fitting scleral lenses, but with the significant advantage of using gas permeable material. These lenses are normally made following the fitting of a final approved PMMA diagnostic lens.

 

 

 

 

 

 

PMMA Scleral Lenses

Scleral lenses made in PMMA can be made to both preformed and impression specifications, as with RGP lenses. Nowadays they are made principally for diagnostic purposes, but they may still have therapeutic and recreational applications.

 

PMMA/RGP Scleral Lenses

PMMA scleral lenses can be made with an RGP optic inserted. Traditionally these have been used where only the 'impression' lens could be successfully fitted, but enhanced oxygenation was required for wear time. They may also be suitable for wearers who have handling difficulties.

 

PMMA Scleral Lens Diagnostic Fitting Sets

The standard fitting set manufactured by Innovative Sclerals Ltd has 9 lenses with 13.5mm scleral radius and 9 lenses with 14.5mm scleral radius. These two ranges are designed with matching progressive apical/limbal clearances to suit most 'normal' and keratoconic eyes likely to be encountered by the practitioner. The lenses are powered within an appropriate range and are unfenestrated for sealed fitting. The fitting set is boxed and supplied with a comprehensive fitting manual. Additional lenses, to specifications outside the normal fitting parameters, can be supplied individually as circumstances arise

 

Fenestration of Scleral Lenses

RGP scleral lenses manufactured by Innovative Sclerals Ltd are designed for sealed fitting with no fenestrations. Occasionally fenestrations or channelling may be required to improve the fitting; in such cases a purpose made PMMA should be tried first to verify the fitting parameters. RGP material, by its nature, is prone to breakage from fenestrations, so they are not normally recommended.

 

Lens Colour

All lenses are supplied clear, with no tints.

 

Lens optic specifications

The final power of lenses is normally determined from the power of the selected diagnostic fitting lens together with any over-refraction and BVD allowance required. Innovative Sclerals Ltd is unable to incorporate cylinder or toric into the lenses.

 

Supply of Products

Innovative Sclerals Ltd will only supply their products to qualified practitioners and appropriate medical departments.

 

Training

Innovative Sclerals Ltd hold regular meetings in the UK. These provide practitioners with theoretical and practical training, irrespective of their previous experience. Other one off meetings can be arranged outside the UK as and when appropriate












RGP lathe turned 23mm
left to right: RGP lathe turned scleral lens 23mm diameter with optic zone decentred 1mm; RGP lathe turned mini-scleral 18mm diameter; RGP scleral lens made from eye impression foreground: standard corneal lens - shown for scale


Single RGP Lathe turned 23mm
RGP lathe turned scleral lens 23mm diameter with optic zone decentred 1mm; left eye marking can be seen at front
















RGP eye impress
RGP scleral lens made from eye impression, approx. 28 x 25mm diameter (this particular lens gives visual improvement and maintains fluid reservoir to war-damaged eye.)